Infrared detector selection for gas analysis and spectroscopy
Developing a process-gas analyser, emission monitor or infrared spectrometer? Start with the measurement method and operating wavelength, then match the detector, amplification, cooling and readout. SMART helps clarify requirements and check candidate configurations with the manufacturer.
NDIR, TDLAS and FTIR: match the method to a detector
Start from the architecture you actually use. These are conditional evaluation routes, not manufacturer approval for a complete gas instrument.
| Method | Define first | Detector checks | Candidate route |
|---|---|---|---|
| NDIR / filtered broadband sensing | Gas absorption band, source, filters, optical path and reference channel. | Check response at the selected band, modulation, background and calibration. | For a single analogue channel, evaluate AM03100-02 or AMS3140-01 after confirming the actual band and signal level. |
| TDLAS / tunable-laser absorption | Laser wavelength and scan, modulation, incident power and interference fringes. | Check response at the laser wavelength, linearity, noise and the required signal bandwidth. | Review PVA-3TE-5 AF for anti-fringing requirements; confirm the full part number and external front end. |
| FTIR / broadband infrared spectroscopy | Required spectral span, beam geometry, throughput and scan speed. | Match the response curve and receiver bandwidth; a cut-off wavelength is not full-system spectral coverage. | Check PVIA-4TE-10 or AMIS8140-01 for the required spectral segment; ask about other configurations if the band extends beyond these models. |
A gas may have several absorption bands. Verify response at the actual wavelength, including for near-infrared lasers. Detector D* does not establish a gas concentration limit. See the manufacturer method references below.
Define the measurement requirements
| Measurement target | Gas / spectral line, concentration and path length |
|---|---|
| Source / architecture | Laser or broadband source; scan and modulation method |
| Optical conditions | Wavelength, spot size, incidence angle and optical power |
| Signal requirements | Bandwidth, noise, dynamic range and readout interface |
| Environment / integration | Ambient temperature, TEC, heatsink and amplifier |
TDLAS detector selection: review fringing and linearity together
- Check response across the actual laser scan, not peak D* alone.
- Laser-induced etalons can involve windows, chips and external parallel surfaces. Review anti-fringing structures, coatings and incidence geometry together.
- Specify optical-power range and linearity. Immersion performance at weak signal levels does not replace saturation checks.
Consider PVA-3TE-5 AF anti-fringing detector selection without assuming system-wide fringe elimination.
Manufacturer references: TDLAS selection · etalon / anti-fringing. These are selection considerations, not SMART test results.
What to confirm before a gas-sensing evaluation
- Light source and gas sample
- Filters and optical coupling
- Detector and amplifier
- Analogue receiver / ADC
- Processing and concentration calibration
- Match source, gas cell / optical path and filters to the target absorption line; include concentration range and interfering gases in the requirements.
- Confirm the analogue front end, output load and ADC or receiver. Detector time constant, module bandwidth and sampling rate are different specifications.
- Check TEC control, external heat dissipation, ambient temperature and supply requirements.
- Agree on background / reference measurement, drift and concentration calibration. The detector does not include a complete gas cell, gas path, source or calibrated analyser.
Products to evaluate
These are starting points for discussion; suitability depends on the requirements above.

AMS3140-01 infrared detection module
2.0–5.5 μm at Tchip −20°C, with 1TE temperature control, differential amplification, 3.3 V supply and a board-to-board interface.

AM03100-02 infrared detection module
1.7–5.9 μm at Tamb 20°C. An uncooled differential-amplifier module with a specified 3.0 V operating point and a 10 × 10 mm solderable PCB.

PVA-3TE-5 AF anti-fringing IR detector
Tunable-laser gas spectroscopy and laser control, selected by spectral line, signal bandwidth and optical configuration.

PVIA-4TE-10 infrared detector
FTIR, longer-wavelength spectroscopy and CO₂ laser measurements within the documented optical-power and linearity limits.
For spectral range, cooling, bandwidth and readout specifications, use the infrared detector selection guide. For multi-point equipment monitoring, continue to industrial sensing requirements.
Manufacturer references and product documents
Reviewed 18 November 2026. Application background does not transfer another instrument’s performance to these components. Use the linked product pages for the manufacturer datasheet and operating conditions.
Start with your measurement requirements
Share the model or application, wavelength, signal bandwidth, quantity and desired schedule. Unknown items can be discussed during selection.
Submit a VIGO inquiry